NEW QUESTION 25
The following values are obtained using a 3-liter calibration syringe during volume calibration of a dry-rolling spirometer:
2.80 L
2.79 L
2.82 L
The performance of this spirometer is best described as

  • A. Neither accurate nor precise.
  • B. Precise, but not accurate.
  • C. Accurate, but not precise.
  • D. Both accurate and precise.

Answer: C

 

NEW QUESTION 26
The following results are obtained:

The RV/TLC ratio from these data is consistent with which of the following?

  • A. Restrictive defect
  • B. Combined obstructive/restrictive defect
  • C. Obstructive defect
  • D. Normal lung volumes

Answer: C

 

NEW QUESTION 27
A pulmonary function technologist is performing an exercise study on a patient with sarcoidosis. Which of the following end-tidal CO2 values should the technologist expect at rest, if the test is performed appropriately?

  • A. 4-5%
  • B. 2-3%
  • C. 0-1.5%
  • D. 7-10%

Answer: A

 

NEW QUESTION 28
When performing quality control in a body plethysmograph using a 5-L isothermal bottle, the VTG at shutter closure are as follows:

A pulmonary function technologist should

  • A. Check biological control before beginning testing.
  • B. Proceed with patient testing.
  • C. Service the mouth pressure transducer.
  • D. Recalibrate the box pressure transducer.

Answer: C

 

NEW QUESTION 29
A 45-year-old female with sarcoidosis has pulmonary function tests annually. The following data have been recorded:

A pulmonary function technologist should conclude that

  • A. The patient's restrictive component has worsened.
  • B. Incorrect predicted values have been used.
  • C. The patient has developed mild obstructive airway disease.
  • D. The patient has a normal rate of loss for FEV1.

Answer: B

 

NEW QUESTION 30
During the calibration and set-up of the metabolic stress testing system for a patient breathing supplemental oxygen, which of the following gas concentrations will ensure accurate calibration of the system?

  • A. Option D
  • B. Option B
  • C. Option A
  • D. Option C

Answer: A

 

NEW QUESTION 31
The following arterial blood gas results are obtained during the final workload of a cardiopulmonary exercise test:

Which of the following best explains these results?

  • A. IV solution has contaminated the blood sample.
  • B. There is air contamination since the PaCO2is so low.
  • C. Blood gas results are normal for someone at end-exercise.
  • D. The test indicates a right-to-left shunt.

Answer: C

 

NEW QUESTION 32
Prior to performing a multiple-breath U2 washout test, a pulmonary function technologist is unable to zero the H2 gas analyzer. Which of the following may be responsible?

  • A. Condensation of water
  • B. Chopper motor failure
  • C. Inadequate vacuum
  • D. Exhausted fuel cell

Answer: A

 

NEW QUESTION 33
During exercise, a subject's oxygen consumption increases out of proportion to his cardiac output. This is due to an increase in

  • A. Coronary blood flow
  • B. Oxygen extraction.
  • C. Anaerobic metabolism
  • D. Alveolar ventilation

Answer: A

 

NEW QUESTION 34
While reviewing exercise data, a pulmonary function technologist notes that the patient's anaerobic threshold occurred at 35% of the patient'sVO2MAX.This indicates the patient has

  • A. cardiac limitation
  • B. given a submaximal effort
  • C. a normal response to exercise
  • D. ventilatory limitation

Answer: A

 

NEW QUESTION 35
While setting up an exercise laboratory in a city with an altitude of 8,600 ft (2,775 m), a pulmonary function technologist notices the fuel cellO2analyzer is displaying 15.2%. Which of the following is the best explanation for this finding?

  • A. F1O2decreases with increasing altitude.
  • B. The analyzer is responding to P1O2.
  • C. The fuel cell needs to be changed.
  • D. This exercise system will not work at high altitude.

Answer: B

 

NEW QUESTION 36
Which of the following tracings represents an airways resistance maneuver performed with a slow breathing frequency?

  • A. Option D
  • B. Option B
  • C. Option A
  • D. Option C

Answer: A

 

NEW QUESTION 37
Using a peak flowmeter, a pulmonary function technologist obtains the following:

The technologist should

  • A. Perform at least one more peak flow trial.
  • B. Report the average of the two best efforts.
  • C. Conclude that bronchodilatation has occurred.
  • D. Report the patient's peak flow as 750 L/min.

Answer: A

 

NEW QUESTION 38
Which of the following is the most reliable method to estimate the effectiveness of standard procedures to minimize the risk of cross-contamination during spirometry testing in a pulmonary function laboratory?

  • A. Observe the handwashing behavior of each technologist.
  • B. Count the number of disposable mouthpieces used for 1 week.
  • C. Ask technologists if they wash their hands before each test
  • D. Ask patients tested about laboratory hygiene.

Answer: A

 

NEW QUESTION 39
While performing a quality control test on an open circuit nitrogen system, the volume of a 3-liter syringe is measured as 3.9 L.
Which of the following is the most probable explanation?

  • A. The initialO2concentration in the syringe was greater than 0.21.
  • B. The volume was not corrected from ATPS to BTPS.
  • C. There was an air leak in the system.
  • D. The nitrogen analyzer gain was set too low.

Answer: C

 

NEW QUESTION 40
A pulmonary function technologist reviews the following home monitoring spirometry results:

According to National Asthma Education and Prevention Program (NAEPP) guidelines, what feedback should the technologist give to the patient regarding test performance?

  • A. Ask the patient to recalibrate the spirometer.
  • B. Ask the patient to blow out longer.
  • C. Encourage the patient to continue testing and monitoring the FEW
  • D. Come into the office for further instructions on proper testing technique.

Answer: C

 

NEW QUESTION 41
A patient who smokes three packs of cigarettes a day has a DLco Of 45% of predicted. The patient is referred for exercise (stress) testing. Which of the following should a pulmonary function technologist select to evaluate oxygenation?

  • A. Arterial blood gas analysis
  • B. End-tidal oxygen tension
  • C. Transcutaneous oxygen electrode
  • D. Pulse oximetry

Answer: A

 

NEW QUESTION 42
Which of the following is an appropriate reason to perform a multiple-breath nitrogen washout test?

  • A. Differentiate obstruction from restriction.
  • B. Measure anatomical dead space.
  • C. Detect early small airway disease.
  • D. Measure oxygen consumption.

Answer: C

Explanation:
Explanation/Reference:

 

NEW QUESTION 43
A helium dilution test has just been performed on a patient. The following results are obtained:
FRC 5.0 L VC 4.0 L ERV 1.5 L
TLC was calculated to be 6.0 L by plethysmography. From this information, a pulmonary function technologist should conclude that the patient

  • A. Did not remain in the helium dilution breathing circuit long enough for equilibration.
  • B. Did not perform the slow vital capacity properly, resulting in too low an FRC by helium dilution.
  • C. Had inadequate intrapulmonary mixing of inspired gas, resulting in an erroneous FRC.
  • D. Was turned into the helium dilution circuit at a lung volume considerably above FR

Answer: C

 

NEW QUESTION 44
......