NEW QUESTION 25
The following values are obtained using a 3-liter calibration syringe during volume calibration of a dry-rolling spirometer:
2.80 L
2.79 L
2.82 L
The performance of this spirometer is best described as
- A. Neither accurate nor precise.
- B. Precise, but not accurate.
- C. Accurate, but not precise.
- D. Both accurate and precise.
Answer: C
NEW QUESTION 26
The following results are obtained:
The RV/TLC ratio from these data is consistent with which of the following?
- A. Restrictive defect
- B. Combined obstructive/restrictive defect
- C. Obstructive defect
- D. Normal lung volumes
Answer: C
NEW QUESTION 27
A pulmonary function technologist is performing an exercise study on a patient with sarcoidosis. Which of the following end-tidal CO2 values should the technologist expect at rest, if the test is performed appropriately?
- A. 4-5%
- B. 2-3%
- C. 0-1.5%
- D. 7-10%
Answer: A
NEW QUESTION 28
When performing quality control in a body plethysmograph using a 5-L isothermal bottle, the VTG at shutter closure are as follows:
A pulmonary function technologist should
- A. Check biological control before beginning testing.
- B. Proceed with patient testing.
- C. Service the mouth pressure transducer.
- D. Recalibrate the box pressure transducer.
Answer: C
NEW QUESTION 29
A 45-year-old female with sarcoidosis has pulmonary function tests annually. The following data have been recorded:
A pulmonary function technologist should conclude that
- A. The patient's restrictive component has worsened.
- B. Incorrect predicted values have been used.
- C. The patient has developed mild obstructive airway disease.
- D. The patient has a normal rate of loss for FEV1.
Answer: B
NEW QUESTION 30
During the calibration and set-up of the metabolic stress testing system for a patient breathing supplemental oxygen, which of the following gas concentrations will ensure accurate calibration of the system?
- A. Option D
- B. Option B
- C. Option A
- D. Option C
Answer: A
NEW QUESTION 31
The following arterial blood gas results are obtained during the final workload of a cardiopulmonary exercise test:
Which of the following best explains these results?
- A. IV solution has contaminated the blood sample.
- B. There is air contamination since the PaCO2is so low.
- C. Blood gas results are normal for someone at end-exercise.
- D. The test indicates a right-to-left shunt.
Answer: C
NEW QUESTION 32
Prior to performing a multiple-breath U2 washout test, a pulmonary function technologist is unable to zero the H2 gas analyzer. Which of the following may be responsible?
- A. Condensation of water
- B. Chopper motor failure
- C. Inadequate vacuum
- D. Exhausted fuel cell
Answer: A
NEW QUESTION 33
During exercise, a subject's oxygen consumption increases out of proportion to his cardiac output. This is due to an increase in
- A. Coronary blood flow
- B. Oxygen extraction.
- C. Anaerobic metabolism
- D. Alveolar ventilation
Answer: A
NEW QUESTION 34
While reviewing exercise data, a pulmonary function technologist notes that the patient's anaerobic threshold occurred at 35% of the patient'sVO2MAX.This indicates the patient has
- A. cardiac limitation
- B. given a submaximal effort
- C. a normal response to exercise
- D. ventilatory limitation
Answer: A
NEW QUESTION 35
While setting up an exercise laboratory in a city with an altitude of 8,600 ft (2,775 m), a pulmonary function technologist notices the fuel cellO2analyzer is displaying 15.2%. Which of the following is the best explanation for this finding?
- A. F1O2decreases with increasing altitude.
- B. The analyzer is responding to P1O2.
- C. The fuel cell needs to be changed.
- D. This exercise system will not work at high altitude.
Answer: B
NEW QUESTION 36
Which of the following tracings represents an airways resistance maneuver performed with a slow breathing frequency?
- A. Option D
- B. Option B
- C. Option A
- D. Option C
Answer: A
NEW QUESTION 37
Using a peak flowmeter, a pulmonary function technologist obtains the following:
The technologist should
- A. Perform at least one more peak flow trial.
- B. Report the average of the two best efforts.
- C. Conclude that bronchodilatation has occurred.
- D. Report the patient's peak flow as 750 L/min.
Answer: A
NEW QUESTION 38
Which of the following is the most reliable method to estimate the effectiveness of standard procedures to minimize the risk of cross-contamination during spirometry testing in a pulmonary function laboratory?
- A. Observe the handwashing behavior of each technologist.
- B. Count the number of disposable mouthpieces used for 1 week.
- C. Ask technologists if they wash their hands before each test
- D. Ask patients tested about laboratory hygiene.
Answer: A
NEW QUESTION 39
While performing a quality control test on an open circuit nitrogen system, the volume of a 3-liter syringe is measured as 3.9 L.
Which of the following is the most probable explanation?
- A. The initialO2concentration in the syringe was greater than 0.21.
- B. The volume was not corrected from ATPS to BTPS.
- C. There was an air leak in the system.
- D. The nitrogen analyzer gain was set too low.
Answer: C
NEW QUESTION 40
A pulmonary function technologist reviews the following home monitoring spirometry results:
According to National Asthma Education and Prevention Program (NAEPP) guidelines, what feedback should the technologist give to the patient regarding test performance?
- A. Ask the patient to recalibrate the spirometer.
- B. Ask the patient to blow out longer.
- C. Encourage the patient to continue testing and monitoring the FEW
- D. Come into the office for further instructions on proper testing technique.
Answer: C
NEW QUESTION 41
A patient who smokes three packs of cigarettes a day has a DLco Of 45% of predicted. The patient is referred for exercise (stress) testing. Which of the following should a pulmonary function technologist select to evaluate oxygenation?
- A. Arterial blood gas analysis
- B. End-tidal oxygen tension
- C. Transcutaneous oxygen electrode
- D. Pulse oximetry
Answer: A
NEW QUESTION 42
Which of the following is an appropriate reason to perform a multiple-breath nitrogen washout test?
- A. Differentiate obstruction from restriction.
- B. Measure anatomical dead space.
- C. Detect early small airway disease.
- D. Measure oxygen consumption.
Answer: C
Explanation:
Explanation/Reference:
NEW QUESTION 43
A helium dilution test has just been performed on a patient. The following results are obtained:
FRC 5.0 L VC 4.0 L ERV 1.5 L
TLC was calculated to be 6.0 L by plethysmography. From this information, a pulmonary function technologist should conclude that the patient
- A. Did not remain in the helium dilution breathing circuit long enough for equilibration.
- B. Did not perform the slow vital capacity properly, resulting in too low an FRC by helium dilution.
- C. Had inadequate intrapulmonary mixing of inspired gas, resulting in an erroneous FRC.
- D. Was turned into the helium dilution circuit at a lung volume considerably above FR
Answer: C